2.1 700 Dosino dosing drive
700 Dosino, Instructions for Use
5
2 Setup
2.1 700 Dosino dosing drive
Checks
Please check immediately on receipt whether the shipment is
complete and undamaged (compare with delivery note and list of
accessories in section 4.4). If transport damage has occurred
please refer to section 4.3 'Warranty'.
Location
The 700 Dosino is a robust instrument and can therefore be used
under rough conditions in laboratories and factories.
However, care must be taken that it is not exposed to a corrosive
atmosphere. Regular care of the instrument is essential if it is
used under rough conditions.
If an instrument which has been stored under cold conditions is
brought into a warm room then it is possible that the atmospheric
humidity could cause water to condense inside the instrument. In
order to avoid damage to the instrument it should be allowed to
acclimatize itself for at least one hour before it is switched on.
2.1.1 Connecting to a control instrument
The 700 Dosino is supplied with one of two different plugs. Please
check that your model is fitted with the correct plug.
•
Model 2.700.0010 with 9-pin plug (DB9)
for connection to the following instruments:
−
726 Titroprocessor
−
729 Dosimat Interface (for connection to 730 Sample
Changer or 726 Titroprocessor)
−
693 VA Processor
−
746 Trace Analyzer
−
695 Autosampler
The
6.2134.010 Adapter cable can be used for connecting Model
2.700.0020 (mini-DIN plug) to the instruments listed above.