2 safety information – Metrohm 700 Dosino User Manual

Page 12

background image

2. Setup

700 Dosino, Instructions for Use

6

Model 2.700.0020 with mini-DIN plug
for connection to the following instruments:

711 Liquino

736, 751, 758, 784, 785 Titrinos

756 KF–Coulometer

The

6.2134.020 Adapter cable can be used for connecting Model

2.700.0010 (DB9-plug) to the instruments listed above.

The location of the correct connection socket for the Dosino can
be found in the ‘Instructions for Use’ of the corresponding control
unit.

Only connect the Dosino to a control instrument when this is
switched off. The control instrument only recognizes the
Dosino during the switching-on process.

Please note the arrangement of the connection sockets. Do
not try to plug in the connection cable by using force!

2.1.2 Safety information

General:
This instrument left our factory in perfect condition from a safety
point of view (see technical data, safety specification). The follow-
ing information must be carefully observed in order to maintain this
condition and to ensure hazard-free operation.

Connection to a control instrument:
This instrument must only be connected to that connection socket
of a Metrohm instrument which is provided for it (see correspond-
ing ‘Instructions for Use’). Only those adapter cables listed in the
instructions are to be used.

Repairs and maintenance:
If faults or malfunctions occur during the operation of the 700
Dosino it is recommended that the connections with the control in-
strument are first checked to see if they are correct.

The Dosino (dosing drive) must not be opened. This should
only be carried out by authorized service personnel.