beautypg.com

3 the dosino in practice, 1 air bubbles are (almost) inevitable – Metrohm 700 Dosino User Manual

Page 29

background image

3.1 Air bubbles are (almost) inevitable

700 Dosino, Instructions for Use

23

3 The Dosino in practice

700 Dosinos can be connected to various control instruments.
They are used as dosing drives which can be operated with differ-
ent 710 Dosing units. The dosing unit forms the buret and is
mounted directly on a reagent bottle or a canister.

Changing a dosing unit (and therefore a reagent) is very easy. The
drive is removed from one dosing unit with a single movement and
attached to the next one. The Dosino remains plugged into the
control instrument.

Before a Dosino buret is used we recommend that the dosing cyl-
inder and the tubing are rinsed through once with reagent solution
in order to ensure that the whole system is bubble-free.
Metrohm instruments possess the comfortable

'PREP' function for

this important preparation step: a complete rinsing cycle is trig-
gered by pressing a key.

10 ml

700 Dosino

10 ml

10 ml

10 ml

10 ml

If a dosing unit is to be filled with a different reagent then the dos-
ing cylinder and connected tubing are first emptied with the
'EMPTY' function. The dosing unit is then filled by using the
'PREP' function .