beautypg.com

Metrohm 700 Dosino User Manual

Page 40

background image

3. The Dosino in practice

700 Dosino, Instructions for Use

34

Dosing unit leaks from distributor

Possible cause:

Leaky distributor disk.

Clean the stopcock washer and distributor disk, see page 29.

Liquid above the piston

Possible cause:

Worn or faulty piston and/or cylinder.

Replace dosing piston and cylinder, see page 13.

Liquid drips into the bottle

Possible cause:

Air in the cylinder.

⇒ Check the tubing ends, in particular that of the aspiration tub-

ing.

⇒ Tighten all tubing connections manually.

Check whether the housing is properly locked. If necessary
remove the housing and then put it back on again.

The reagent releases a lot of gas, i.e. dissolved air forms
bubbles.

⇒ Carry out PREP.
⇒ Lower the filling speed.

If necessary degas the reagent, i. e. with helium, vacuum or
an ultrasonic bath.

Wear and tear.

Replace piston and/or cylinder.

PREP not carried out or incorrect parameters.

⇒ Carry out PREP.

Correct tubing lengths and diameters.

Dosing unit is not recognized or incorrectly recognized

Possible cause:

Dosing drive not attached correctly.

⇒ Check whether the nominal volume on the dosing unit housing

is correct.

⇒ Remove Dosino and put it back on again.
⇒ Check whether the drive is correctly seated.
⇒ Switch control instrument off and on again.

If necessary contact your local Metrohm service center.