5.1 Technical specifications
748 DH Sample Changer
20
5 Appendix
5.1 Technical specifications
Lift
Max. lift way
146 mm
Max. load
500 g
Sample position
precision
±
2 mm
Sample racks
48 beaker, 250 mL
136 beaker, 75 mL
Material
Light metal, powder coated
Pump
Capacity
0.33 L/min
Fuse
for 100...120 V: T1.25A
for 220...240 V: T630mA L
Housing
Al (antichromed) and stainless steel, powder coated
RS232 interface
for PC connection
Input/output lines
for beaker sensor and any other devices
Ambient temperature
Nom. operation range
0 ... 45 °C
Storage, transport
– 25 ... 60 °C
Safety specifications
Designed and tested in accordance to IEC publication
1010-1, safety class I. This manual contains information
and warnings which have to be followed by the user to en-
sure safe operation and to retain the apparatus in safe
condition.
Mains connection
Voltage
100, 120, 220, 240 V
±
10 % (switchable)
Frequency
50 ... 60 Hz
Power consumption
max. 190 VA
Fuses (2)
for 100...120 V: T2A/120V
for 220...240 V: T1A/250V
Dimensions
Width
750 mm
Height
750 mm
Depth
600 mm
Weight, incl. accessories
ca. 35.5kg