1.6 Multilinear Regression
NIRS Vision Tutorial – Operation
19
Perform Sample Selection
1.
Click on Select Apply Math.
2.
In the Edit Sample Selection Parameters dialog box enter the
following settings:
Selection Method Mahalanobis Distance in Principal Component
Space
Math Treatment 2nd derivative
Leave all other settings unchanged
3.
Click on [OK].
4.
In the Edit Math Properties dialog box in the Principal Components
ID Parameters tab set the following wavelength regions:
Region 1
Minimum: 416
Maximum: 1080
Region 2
Minimum: 1120
Maximum: 2484
Cum Var. 95 %
Click on [OK]
5.
In the Edit Math Properties dialog box in the 2nd derivative tab set
the following values:
Segment: 10
Gap:
0
6.
In the Sample Selection Options dialog box apply or confirm the
following settings:
Find Outliers
enabled
Outlier Threshold
0.95
Threshold Type
Probability Level
Find Redundant
enabled
By number of samples
enabled
Training Set
75 %
Acceptance Set
25 %
Click on [OK]
Vision now performs calculations according to the parameters defined in the
previous steps.