Click on the + sign next to the green cuvette icon, Click on “acdisol, Click on select ( apply math – Metrohm Vision – Tutorial User Manual
Page 17

4 Qualitative Model Development
10
NIRS Vision Tutorial – Operation
Open Project “xdsstart”
1.
Click on the + sign next to the green cuvette icon.
Add Product to the Library
1.
Drag the product “AcDiSol” from the products folder into the library
folder.
2.
In the Library Product dialog box confirm the product ID and click on
[OK].
The product “AcDiSol” now appears under the library name next to a gray
(empty) cuvette. This indicates that “AcDiSol” has been chosen for sample
selection but has not yet completed the selection process.
Start Sample Selection
1.
Click on “AcDiSol”
2.
Click on Select Apply Math
3.
In the Edit Sample Selection Parameters dialog box, enter the
following information:
Selected Method “Mahalanobis Distance in Principal Component
Space”
Math Treatment “2nd derivative”
Click on [OK]
4.
In the Edit Math Properties dialog box in the Principal Components
ID Parameters tab set the following wavelength regions:
Region 1
Minimum:
416
Maximum:
1080
Region 2
Minimum:
1120
Maximum:
2484
Cum Var.
95 %
Click on [OK]
5.
In the Sample Selection Options dialog box, set the following:
Find Outliers
enabled
Outlier Threshold
0.95
Threshold Type
Probability Level
Find Redundant
enabled
By number of samples
enabled
Training Set
75 %
Acceptance Set
25 %
Click on [OK]