Click on identify ( save results, 4 qualifying a product, Click on select ( apply math – Metrohm Vision – Tutorial User Manual
Page 21: Qualifying a product

4 Qualitative Model Development
14
NIRS Vision Tutorial – Operation
Save Results
1.
Click on Identify Save Results
2.
If asked “Save identify method #1 as library global identify
method?” click on [YES]
1.4
Qualifying a Product
The 3 stearate products are ambiguous.
A Qualification Method is used to determine how similar a sample is it to what
has been accepted in the past. It can also be used to distinguish very similar
products, e. g. Stearates.
Create a Qualification Method
1.
Click on Mode Qualitative Analysis Qualify Method
Development.
The currently selected library appears next to a red cuvette in the
Samples View.
2.
Double click on the xdslib library
all products are listed under the selected library.
3.
Double-click on the “Calcium Stearate”
The cuvette turns gray, indicating that the product has been selected.
NOTE: the cuvette must turn gray before the math can be applied.
4.
Click on Select Apply Math.
5.
In the Edit Qualify Method dialog box apply or confirm the following
settings:
Method
Maximum Distance in Wavelength Space
Pre Treatment Method
2nd derivative
Threshold
type = Match value
value = 3
Library Stabilization
1
Click on [OK]
6.
In the Edit Math Properties dialog box in the wavelength distance tab
set the following wavelength regions:
Region 1
Minimum: 1870
Maximum: 2000
Click on [OK]