beautypg.com

Customer service – Bird Technologies 4308 User Manual

Page 30

background image

I

Make all connections as short as possible to avoid mismatches and addi-
tional insertion losses.

I

Make all calibration checks on each range at the upper one-third of the
scales.

I

If the unit is found to be outside the stated accuracy, the unit must be re-
turned to the factory for recalibration.

Customer Service

Any maintenance or service procedure beyond scope of those provided in this section
should be referred to a qualified service center. Bird Electronic Corporation maintains
complete repair and calibration facilities at the following address:

Standard

Element

RF Power Source

Coaxial

Cable

Model 43

Filter

Load

Power Meter

Power Head

Model 4308
(Wattmeter
Under Test)

440-96

0 MH

z

4308-1

50W

1.5

5

15

50

TRANSIT

50W

50E

400-1

000 MHz

Figure 7

Calibration Check Set-up

Bird Model 4308 Thruline Wattmeter

&