There are no provisions for mounting of these units. These units are designed for
use in a vertical position, however, if the situation should arise, they may be
used in any position. If the unit is to be used in any other position, be sure the
mounting is substantial and clearance around the load is maintained for cooling.
Placement
For ease of movement, hold the two handles on each side of the RF input con-
nector, located at the top center of the unit. Move the unit into position, and
rest it on the four bumper feet.
Interlock Connections
There are two terminals on the interlock connection of the load, with a rating of
10A @ 120V AC and 5A @ 250VAC. Check the requirements of the transmitter
interlock and make the connections to the appropriate terminals as required.
•
Attach the transmitter interlock connections to the two binding posts
on the control panel. See Figure 2.
Figure 2 Interlock Terminal
CAUTION
Do not block air flow. Air enters through perforated grilles at the top of the
unit and exhausts through a lower grill in the unit. Blocking these grilles
could cause unit failure.
WARNING
Although the interlock terminal is rated to 230 VAC, limit the interlock terminal
to less than 30 Vrms. Otherwise, the potential for electrical shock exists.
CAUTION
Ensure the 115/230 voltage selector on the front panel of the unit is set to
the proper voltage before AC power is applied. Incorrect voltage setting can
permentaly damage the unit.