Identification of Parts

Operation System

(3)Flow of cooling water

●In distillation of boiler water ,cooling water flows and drains from the

decompressor,

cooling

water solenoid valve to

condenser. When distilled water tank is full of water or taking ion-

exchanged water , distillation stops and water flow and drain stops automatically.

(4)Collection of distilled water

● For about 10mim. from the beginning of distillation of boiler water , distilled water condensed in

condenser is stored into

distilled water tank through

electrode for distilled water, after

initial water is drained by

solenoid valve (to drain the initial distilled water) opened. The unit judges

water full and stops distillation when the highest

float switch inside the tank is activated.

Distilled water stored is taken out by

pump(to take distilled water) through

solenoid valve to

take distilled water,

distilled water flow sensor (WG221S only)and

distilled water outlet.