Maintenance

When you suspect “Trouble?”

Trouble & Countermeasure

Check the following points if there should or seem to be some machine trouble.

Trouble

Cause

Countermeasure

The flask wouldn’t rotate

● Volume is at the “min”

● Disconnection of motor cable

● Incomplete set-up or fastening

of rotary joint cause racing

● Something touches the flask

● Turn the volume knob up

● Insert into the socket on the

controller

● Fasten the rotary joint holder and

re-set

●remove something that contacts

Incomplete vacuumization

● Wear and deterioration of

vacuum seal

● Cooler nut is incompletely

fastened

● Glass apparatus break

● Incomplete connection of glass

apparatus

● Leak from pipe joint

● Exchange of vacuum seal

● Re-fasten

● Exchange

● Put vacuum grease on

●Stop the leak

if you have any question, contact the seller or Yamato Scientific office.