1

Procedure
Note: the procedure below shows calibration for the commonly used
-0 mA format. If you have set the outputs to any other format, the
unit will prompt you with the output high and low values you have
chosen.

When the

display shows

Action/Description

Put the instrument in setup mode and scroll through the main menu

CALOut_n

Press

ì or Å

CAlOut_y

Press

PGM to select output calibration

Out_LO_=

Connect the multimeter to measure the output level, then press

PGM

4.00

Press

PGM

Measure the output signal

Adjust the output (using the

ì or Å keys) until the output is at the value shown

When you are happy that the output is correct, press

PGM

Out_HI_=

Press

PGM

20.00

Press

PGM

Measure the output signal

Adjust the output (using the

ì or Å keys) until the output is at the value shown

When you are happy that the output is correct, press

PGM

Save

Continue with the setup sequence.

Note: Do not remove the power while the save message is on display.