Focused Lamp Installation Instructions
1. Turn Power Off.
2. Disconnect all power and high voltage cable to irradiator cover.
3. Remove lamp irradiator cover by removing the (2) 10-32 knobs at each end of the irradiator.
4. Place the irradiator upside down on its handles.
5. Refer to drawing 1874 sheet 1 of 4 (see Diagrams section).
6. NOTE: mertal air diffuser use only with 4, 6 lamp assemblies and laboratory conveyorized system.
A. Remove one metal diffuser. Loosen the 8-32 screws on the lampholder to release the top bracket.
Do not remove the bracket--this will swing away.
Clean the UV lamp and reflectors with a lint free cloth and denatured isopropyl alcohol (we recom-
mend the use of disposable latex or vinyl gloves to keep th elamp free of ingerprints which will burn
into the lamp or liners and create hotspots, disallowing the transmission of the UV rays).
B. Place one wire and ceramic end of the lamp through the mounted metal air diffuser and on top of the
lampholder. Place the loose metal air diffuser around the other lamp end and place the lamp end
onto the lampholder.
BC-300 Benchtop UV Curing System
5
6. Note: metal air diffuser used only with 4, 6 lamp assemblies and laboratory conveyorized system.
A. Remove one metal air diffuser.
Clean the UV lamp and reflectors with a lint free cloth and denatured isopropyl alcohol (we recom-
mend the use of disposable latex or vinyl gloves to keep the lamp free of fingerprints which will burn
into the lamp or liners and create hot spots, disallowing the transmission of the UV rays.)
B. Place one wire & ceramic end of the lamp through the mounted metal air diffuser and on top of the
lampholder. Place the loose metal air diffuser around the other lamp end and place the lamp end
onto the lampholder.
C. Remount metal air diffuser with 8-32 x 3/16 screws. Remove the non-focused bracket and the 8/32
x 1/2 screws from the package. At one end feed the wire ceramic end through the lager hole in the
non-focus bracket. Mount non-focus bracket to the lampholder with the 8/32 x 1/2 screws provided.
Mount the wire lead from the lamp end to the ceramic stand off and replace the 8-32 nut. Repeat on
opposite end.
7. Place the irradiator back in place and tighten the (4) 10-32 knobs. Reconnect the high voltage cable by
matching index insuring proper connection and screw connector cable down completely.
CAUTION: when fixture is operating, ultraviolet rays are emitted which are harmful to eyes and skin. Great
care should be taken to insure personnel are not exposed to direct or reflected radiation. Suitable eye and
skin protection, such as uvps uv blocking spectacles or goggles should be worn when lamp is in operation.
Before cleaning or relamping, ALWAYS TURN POWER OFF.