UVP Benchtop Curing System User Manual
Page 5
Lamp Installation and Relamp Instructions
A drawing illustrating the lamp installation is included under the Diagrams section (see sheet 1874 1 of 4).
1. Turn power OFF.
2. Disconnect Hi-Voltage cable from irradiator cover. Refer to Wiring Diagram drawing number 1871.
3. Remove lamp irradiator cover by removing the 2 screws at each and of irradiator.
4. Place irradiator on its handles.
5. Remove top lampholder bracket from each side of lamp.
6. Side opposite the connector, remove aluminum cover and ceramic washer.
7. Slide lamp out.
8. When installing lamp, slide lamp between ceramic washer and aluminum holder. Use gloves to prevent
fingerprints on lamp.
9. Replace aluminum holder and ceramic washer to assembly.
10. Center lamp so lamp ends are equally spaced on large lampholder bracket.
11. Replace top lampholder brackets, tightening each side equally. Clean lamp a with dry cloth and alcohol.
12. Replace assembly into side extrusion and secure with the 4 screws.
13. Re-connect Hi-Voltage connector.
14. Return to Step 3 of General Instructions.
CAUTION: When fixture is operating, ultraviolet rays are emitted which are harmful to eyes and skin. Great
care should be taken to insure personnel are not exposed to direct or reflected radiation. Suitable eye and
skin protection, such as UVPs UV Blocking Spectacles or Goggles should be employed when lamp is in
operation. BEFORE CLEANING OR RELAMPING, TURN POWER OFF.
Non-Focused Lamp Installation Instructions
A drawing illustrating the lamp installation is included under the Diagrams section (see sheet 1874 3 of 4).
1. Turn power off.
2. Disconnect all power and high voltage cable to irradiator cover.
3. Remove lamp irradiator cover by removing the (2) 10-32 knobs at each end of the irradiator.
4. Place the irradiator upside down on its handles.
5. Refer to drawing 1874 sheet 3 of 4 (see Diagrams section).