UVP HybriCycler User Manual
Page 6

HC-3000 HybriCycler Hybridization Oven
6
3. Press the “Motor” switch to the ON
position.
Setting the Temperature Controls
Whenever power is applied to the controller, the software revision number is displayed
for several seconds. While the software revision number is being displayed, the intensity
of the display digits alternates between full and half brightness, and on some models the
two outside decimals points blink on and off. After several seconds, the display reverts
to showing the controlled temperature. In this mode, at most a single decimal point is
illuminated, and the display intensity is steady.
Altering the Setpoint
The current setpoint value can be altered using the “UP” and “DOWN” buttons while the
setpoint is being displayed. To change the setpoint from normal mode, proceed as
follows:
• Momentarily press then release either the “UP” or the “DOWN” button. The LED
intensity will alternate between half and full intensity to indicate that the displayed
value is the current setpoint.
• Increase or decrease the setpoint value by pressing “UP” or “DOWN” buttons
respectively. If either button is held down for more than several seconds, setpoint
value will increase or decrease continuously.
• When the desired setpoint is reached, wait approximately five seconds without
pressing either button; the display will revert to normal mode showing the actual
temperature.
The new setpoint becomes effective and is stored in the non-volatile memory when the
display reverts to the normal mode.
Calibrating the Temperature Reading
The HybriCycler is calibrated at the UVP factory. UVP recommends temperature
recalibration be performed at the UVP factory during the warranty period as recalibration
by the user may void the warranty. If the user chooses to recalibrate the unit,
Temperature Calibration Reading instructions may be obtained from UVP by calling
UVP’s customer service department in Upland, California at (800) 452-6788 or (909)
946-3197 or Cambridge, UK at +44(0)1223-420022.