Limited warranty – SATEC EM720 Installation Manual User Manual
Page 2

ii
LIMITED WARRANTY
The manufacturer offers the customer a 24-month functional warranty on the instrument for faulty
workmanship or parts from date of dispatch from the distributor. In all cases, this warranty is valid for 36
months from the date of production. This warranty is on a return to factory basis.
The manufacturer does not accept liability for any damage caused by instrument malfunction. The
manufacturer accepts no responsibility for the suitability of the instrument to the application for which it
was purchased.
Failure to install, set up or operate the instrument according to the instructions herein will void the
warranty.
Only a duly authorized representative of the manufacturer may open your instrument. The unit should
only be opened in a fully anti-static environment. Failure to do so may damage the electronic
components and will void the warranty.
NOTE
The greatest care has been taken to manufacture and calibrate your instrument. However, these
instructions do not cover all possible contingencies that may arise during installation, operation or
maintenance, and all details and variations of this equipment are not covered by these instructions.
For additional information regarding installation, operation or maintenance of this instrument, contact
the manufacturer or your local representative or distributor.
IMPORTANT
Please read the instructions in this manual before performing installation, and take note of the following
precautions:
1. Ensure that all incoming AC power and other power sources are turned OFF before performing any
work on the instrument. Failure to do so may result in serious or even fatal injury and/or equipment
damage.
2.
Before connecting the instrument to the power source, check the labels on the front of the instrument to
ensure that your instrument is equipped with the appropriate power supply voltage, input voltages and
currents for your application.
3.
Do not connect the instrument to a power source if it is damaged.
4.
Do not expose the instrument to rain or moisture.
5.
The secondary of an external current transformer must never be allowed to be open circuit when the
primary is energized. An open circuit can cause high voltages, possibly resulting in equipment damage,
fire and even serious or fatal injury. Ensure that the current transformer wiring is made through shorting
switches and is secured using an external strain relief to reduce mechanical strain on the screw
terminals, if necessary.
6.
Only qualified personnel familiar with the instrument and its associated electrical equipment must
perform setup procedures.
7.
DO NOT open the instrument under any circumstances.
)
Read this manual thoroughly before connecting the meter to the current carrying circuits. During
operation of the meter, hazardous voltages are present on input terminals. Failure to observe
precautions can result in serious or even fatal injury or damage to equipment.