50 • Calibrating the Model GX-2001 with the Calibration Station

9. To begin calibration, click the “Cal. Unit Calibration” button on

the Download window.

10.After the Model GX-2001 has been calibrated, remove it

from the Charger/Downloading Station by pressing down on
the metal release lever located on the right side of the
Calibration Adapter.

11.Place another Model GX-2001 in the Data Logging/Charger

Station if you have other unit(s) to calibrate. Repeat steps 5 –
10 above.

12.When you are done calibrating, remove the regulator from

the gas cylinder.

13.Disconnect the regulator from the Span Gas In tubing. Store

the regulator and gas cylinder in a safe place.

14.Turn off the Calibration Station.