Calibrating the Model GX-2001 with the Calibration Station • 49
Manual Calibration
If the Model GX-2001 is not due for calibration, you can still
calibrate the unit with the Calibration Station using the Cal. Unit
Calibration button on the Download window.
NOTE: If manually calibrating a unit more than once in an hour,
see the Calibration Interval section below for
precautions.
1. Make sure you have properly setup the GX-2001 Calibration
Station following the instructions under “Preparing for
Calibration.”
2. Turn on the computer, then launch the GX-2001 Data
Logging Software.
3. Make sure that the Automatic Calibration box on the
Download window is unchecked, or left blank.
4. Using the On/Off switch, turn on the GX-2001 Calibration
Station.
5. Insert the Model GX-2001 into the Charger/Downloading
Station on top of the GX-2001 Calibration Station.
6. Lock the Calibration Adapter onto the Model GX-2001 into
place by pressing the black button on the right side of the
Calibration Adapter. (Refer to Figure 26.)
7. Turn on the Model GX-2001 by pressing and holding the
Mode button on the Model GX-2001 for about one second.
8. Download information from the Model GX-2001 by clicking
either the Instrument Information button or the Complete
Download button. This action activates the Cal. Unit
Calibration button.