RKI Instruments Eagle Series User Manual
Page 38

Diffusion Eagle Instruction Manual
Setup Mode • 33
Updating the full-scale setting
1. Press the SHIFT/
▼
button to place the prompt in the second line, then
press the POWER/ENTER button to update the full-scale setting. The
full-scale setting flashes.
The maximum full-scale setting for the oxygen channel is 40.0 VOL%;
the minimum setting is 25.0 VOL%. The default setting is 40.0 VOL%.
2. Press the AIR/
▲
and SHIFT/
▼
buttons to display the desired full-scale
setting, then press the POWER/ENTER button to enter the setting. The
prompt flashes.
Updating the display increment setting
1. Press the SHIFT/
▼
button to place the prompt in the third line, then press
the POWER/ENTER button. The display increment setting flashes. The
allowable settings are 0.2 VOL% (default) and 0.5 VOL%.
2. Press the AIR/
▲
or SHIFT/
▼
button to display the desired display
increment setting, then press the POWER/ENTER button to enter the
setting. The prompt flashes.
Returning to the main menu
1. Press the SHIFT/
▼
button. The ESCAPE message displays. (Press the
AIR/
▲
button to return to the previous screen.)
2. Press the POWER/ENTER button to save the new settings. The OTHER
GAS SET message displays, then the Gas Combinations menu displays.
3. To exit the Gas Combinations menu, press the SHIFT/
▼
button until the
prompt is next to Channel 4, then press the SHIFT/
▼
button again. The
ESCAPE message displays.
4. Press the POWER/ENTER button. The message SAVING DATA
displays, then the main menu displays.
Updating toxics channel settings
This section describes how to update the target gas label, set a custom gas
label, and update the full-scale and display increment settings for a toxic gas
channel.