32 • Setup Mode
Diffusion Eagle Instruction Manual
Returning to the main menu
1. Press the SHIFT/
▼
button. The ESCAPE message displays. (Press the
AIR/
▲
button to return to the previous screen.)
2. Press the POWER/ENTER button to save the new setting. The OTHER
GAS SET message displays, then the Gas Combinations menu displays.
3. To exit the Gas Combinations menu, press the SHIFT/
▼
button until the
prompt is next to Channel 4, then press SHIFT/
▼
again. The ESCAPE
message displays.
4. Press the POWER/ENTER button. The message SAVING DATA
displays, then the main menu displays.
Updating oxygen channel settings
This section describes how to update the target gas label, full-scale setting,
and display increment setting for the oxygen channel.
Updating the target gas label
1. From the main menu, select the GAS COMBINATIONS menu option.
2. Use the AIR/
▲
or SHIFT/
▼
button to place the prompt next to the oxygen
channel (in this example OXY).
3. Press the POWER/ENTER button. The oxygen target gas label flashes.
This indicates that this setting can now be updated.
4. Press the AIR/
▲
or SHIFT/
▼
buttons to display the available oxygen
target gas labels (OXY, ***, and NOT USED).
CAUTION:
The *** setting is not intended for customer setup.
Contact RKI Instruments, Inc., before using this setting
for the oxygen channel.
5. Press the POWER/ENTER button to enter the new target gas label.
6. To exit the Gas Combinations menu, press the SHIFT/
▼
button until the
prompt is next to Channel 4, then press the SHIFT/
▼
button again. The
ESCAPE message displays.
7. Press the POWER/ENTER button. The SAVING DATA message
displays, then the main menu displays.