RKI Instruments Eagle Series User Manual
Page 36

Diffusion Eagle Instruction Manual
Setup Mode • 31
3. Press the POWER/ENTER button to enter the displayed character. The
next character flashes.
4. Repeat steps 2 and 3 to enter the remaining characters. When you enter
the last character, the prompt flashes.
Updating the full-scale PPM setting
CAUTION:
The full-scale PPM setting must correspond to 100% LEL
for the target gas in order for the Eagle to display
accurate PPM readings for the combustible gas channel.
1. Press the SHIFT/
▼
button to place the prompt in the second line, then
press the POWER/ENTER button to update the full-scale setting. The
full-scale setting flashes.
The maximum full-scale setting for the combustible gas channel is
50,000 PPM; the minimum setting is 1000 ppm. The default setting is
50,000 ppm.
2. Press the AIR/
▲
and SHIFT/
▼
buttons to display the desired full-scale
setting (see Table 5), then press the POWER/ENTER button to enter the
setting. The prompt flashes.
Table 5: full-scale PPM Readings
Equivalent to 100% LEL
Target Gas
full-scale Setting
Methane (CH
4
)
50,000 ppm
Hexane
11,000 ppm
Hydrogen
40,000 ppm
Pentane
15,000 ppm
Styrene
9,000 ppm
IPA
20,000 ppm
Isobutane
18,000 ppm
Propane
21,000 ppm
Propylene
20,000 ppm
Toluene
11,000 ppm
Ethane
30,000 ppm
Ethanol
33,000 ppm
Benzene
12,000 ppm