beautypg.com

Practical Instrument Electronics 830 User Manual

Page 3

background image

Page 1

General Information

Get more tools in a smaller calibrator

Carry eight single function calibrators

plus

a milliamp calibrator with loop supply

plus

a loop troubleshooter in the palm of your hand!

Milliamp • Voltage • Frequency • pH • Ohms

Thermocouples • RTDs • Pressure

Loop Diagnostics • Transmitter Supply

1

Detect ‘hidden’ loop problems

Quickly diagnose troublesome ground faults & current leak-
age with patented Loop Diagnostic technology. These prob-
lems are undetectable with other instruments!

2

Check all loop parameters at once with LoopScope™
Simultaneously displays current, voltage and resistance
to let you know the condition of a live loop. Finds prob-
lems with power supplies & loops with too many loads.

Patented by PIE!

3

Automatically detect 2, 3, or 4 wire RTDs

Trouble shoot sensor connections and find broken wires with
patented technology. LCD indicates which of the four wires
are connected to an RTD sensor.

4

Swap out faulty transmitters to diagnose control issues

Easily setup the 830 as a universal isolated transmitter for
Thermocouple, RTD, frequency, millivolt, resistance or pres-
sure. If the loop is back under control you know the

transmitter was faulty. Turn on the LoopScope to

simultaneously see the loop current, loop voltage and loop
loads for troubleshooting.

5

Troubleshoot wiring problems without a multimeter

Built in continuity checker with ‘beeper’ quickly finds broken
wires or shorts in instrumentation wiring. Also handy for
checking operation of relays and controller outputs.