Page 12

Calibrate a 2-Wire Transmitter by sourcing the input while
monitoring the output.

Works with SOURCE pH, T/C, DC V, OHMS, RTD, FREQ and
READ PRESSURE.

Move the power switch

w

to READ and Double click the

e

DIAL KNOB and the MAIN menu for the function in use will
appear for 15 seconds:

Turn the

e

DIAL KNOB to move to the second (or third) menu

page so the word

mA DISPLAY appears at the top of the

menu.

Turn the

e

DIAL KNOB to move through the menu. Press the

e

DIAL KNOB to toggle between OFF and ON or to change the
MODE setting.

HART 250

Ω - turn on the 250Ω resistor if you are powering up

a HART transmitter. This provides the loop load required for
HART communication.

EXIT MENU - exits this menu immediately and saves any
changes. Menu will automatically exit after 15 seconds of
inactivity.

mA DISPLAY

> EXIT

MODE

OFF

READ PWRM READ% PWRM%

HART 250

Ω OFF

ON

LEAK DETECT OFF

ON

MAIN

> EXIT

FUNCTION

T/C

UNITS

°C

T/C TYPE

K

COLD JUNC

ON