Controller Specifications

System Component Specifications

Controller Specifications

Environmental Requirements—Series 494 Hardware

All Series 494 hardware components are intended for indoor use only. This
indoor environment must conform to the following environmental specifications.

Note

All Series 494 Controller must only be operated under the installation
and ambient conditions (such as, temperature, moisture, and EMC)
specified.

Note

To maintain EMC compliance, the controller must be installed in a
location that does not exceed the EN 61000-6-4 emission standard for
industrial environments.

Parameter

Specification

Temperature

5ºC–40ºC (41ºF–104ºF)

Humidity

5–85%, non-condensing

Altitude

3048 m (10,000 ft) maximum

Space
Requirements

For proper ventilation, allow 51 mm (2 in) clearance
on all sides of the chassis.

The rear of the chassis requires a minimum
clearance of 15.24 cm (6 in) for cable connections.