beautypg.com

Model 505.20/.30 specifications, Model 505.20/.30 specifications 20, Model – MTS Landmark Site Prep Guide User Manual

Page 20: 20/.30 specifications

background image

Landmark Test System Site Prep Guide

20

Hydraulic Power Unit Specifications

System Component Specifications

Model 505.20/.30 Specifications

General Specifications

Parameter

Specification

Environmental

Operating temperature

Humidity

Heat load

Noise

*

rating at 1 m

*

Sound pressure level [db (A)] is expressed as a free field value. Readings may vary with the
acoustic environment.

For use in a controlled environment.

5–40°C (41–104°F)

0–85% noncondensing

< 630 kcal/hr (2500 Btu/hr)

63 dB (A) fully compensated

Hydraulic fluid

Typical operating
temperature range

Filtration (microns)

Reservoir capacity

Exxon Mobil DTE 25 or Shell Tellus 46 AW

43–49°C (110–120°F)

3 microns nominal

341 L (90 gal) maximum
208 L (55 gal) minimum

Heat exchanger

Flow requirements, cooling

Water pressure

Maximum pressure

Water hose

Water-cooled

See Water flow rating (input temperature)

0.24–0.31 MPa (35–45 psi differential)

0.83 MPa (120 psi)

25 mm (1 in)—inside diameter

Hydraulic connections

Pressure

Return

Drain

Each requires an O-ring face seal

-16

-16

-6 and -8 (two connections)

Water connections

-16 (1 in NPT) for both In and Out

Electrical power

Line voltage

Control Voltage

200-575 V AC, 3

Φ

at 50/60 Hz

24 V DC

Dimensions

Height

Length

Width

1423 mm (56 in)

1575 mm (62 in)

864 mm (34 in)