Fill the pressure intensifier, Fill the pressure intensifier 60 – MTS Model 286-31 Transient Permeability Pore Pressure Intensifier User Manual
Page 60
286.31 Transient Permeability Pore Pressure Intensifier
60
Running a Transient Permeability Flow Test
Operation
Fill the Pressure Intensifier
1. If you have a specimen installed in the triaxial cell, remove it now. Your
triaxial cell manual has the details on doing this.
2. Turn the valves to match the positions shown in the Step 3 picture.
3. Fill the Fluid Reservoir. Skip this step if the reservoir is 3/4 full.
If your 286.31 is connected to a distilled water supply —
Turn the Reservoir Fill/Drain valve to the Open position.
When the Fluid Reservoir is about 3/4 full, turn the Reservoir Fill/
Drain valve to the Close position.
If your 286.31 is not connected to a distilled water supply —
Remove the filler plug at the top of the 286.31.
Open the console’s rear door. Unscrew one of the hex–head plugs in the
Reservoir. (This allows air to vent more easily out of the Reservoir.)
Use a funnel to fill the Fluid Reservoir about 3/4 full.
Seal the filler cap and hex–head plug with Teflon thread sealing tape
and reinstall them.