286.31 Transient Permeability Pore Pressure Intensifier
6
Major Components
Introduction
Major Components
1
vent
is a removable cap that allows air trapped in the system to escape; also the
fill point for manually filling the reservoir.
2
cabinet
houses the 286.31. It has leveling casters and a rear access door.
3
fluid reservoir
holds the system’s distilled water in a clear acrylic container.
4
control panel
has the valves needed to operate the 286.31. Valves are identified by name
and placed in front of a hydraulic schematic of the 286.31 and Triaxial Cell.
5
identification label
lists unit specifications such as pressure, reservoir capacity, and fluid type.
6
pressure intensifier
is an servovalve controlled actuator that operates a pump which boosts pore
pressure fluid up to 10,000 psi (69 MPa).
7
actuator manifold
is the connection point for hydraulic components.
8
servovalve
responds to system control signals to operate the pressure intensifier’s
piston.
9
reference volumes
are two high pressure reservoirs used during transient permeability tests.
1
2
3
4
5
6
7
8
9