MTS Model 286-30 Pore Pressure Intensifier User Manual
Page 16

286.30 Pressure Intensifier
16
Description of Major Components
Introduction
Major Components
Item
Component
Description
1
Cabinet
Enclosure mounted on casters. Includes leveling legs for stability and a
hinged rear access door.
2
Reservoir
Clear acrylic container. Includes filler port with strainer which also acts as a
vent for escaping gases.
3
Operator control
panel
The panel contains all the valves used during operation and includes a
schematic of the unit as an aid during operation.
4
Pressure
transducer
Provides a signal which represents the level of pore fluid pressure being
applied to the specimen. The signal is used by the electronic controller
equipment as feedback when pressure is the controlled variable.
5
Pressure relief
valve
Relieves pressure if pressure in the system exceeds the PPI’s pressure rating
by a certain amount.
6
Pressure
intensifier
Actuator which uses servovalve controlled hydraulic fluid to pressurize the
pore fluid. The top end, identified as “6A,” is the high pressure, pore fluid
part of the actuator. The lower part, “6B” is the low pressure, hydraulic fluid
end.
7
Hydraulic service
manifold
Provides a hydraulic interface between the external hydraulic power supply
and the pressure intensifier. Provides mounting for various components
such as the servovalve and accumulators.
8
Servovalve
Controls flow and pressure of hydraulic fluid applied to the input side of the
pressure intensifier.
9
Linear transducer
Provides a signal that is proportional to the linear displacement of the
pressure intensifier’s piston relative to a fixed reference.