Installing a Collet for a Round Specimen

To install a collet assembly for a round specimen into the grip housing:

Round Collet

1. Ensure that the grips are in the release condition and that hydraulic pressure

is turned off.

2. Clean the specimen contacting surfaces of the collet with acetone or any

other volatile solvent that does not leave a residue.

3. Apply a thin layer of Molykote G·n paste (part of the grip service kit) to the

threaded surfaces of the collets.

4. Use a spanner wrench to thread the collet assembly into the grip housing

until the top surface of the collet is flush with the top surface of the grip
housing.

5. Adjust the collet assembly as necessary until the alignment mark on the

collet corresponds to the alignment mark on the grip housing. Use a hex key
wrench to thread the socket head cap screw with bumper into the threaded
hole in the grip piston.

Top View

Cutaway
Side View

Spanner
Wrench Holes

Alignment
Mark

Coat the threads on
the collet with
Molykote G-n paste

Bumper

Collet Retainer

Socket Head
Cap Screw