Function, Adjustment options, Atmocontrol fda version – Memmert AtmoCONTROL FDA Version 2.X.X User Manual
Page 20: Change humid- ity, Name of programme segment, Duration, Target (setpoint) humidity, Tolerance value above/below, Spwt, Hold pres- sure

20
D30396 | Date 09/2014
AtmoCONTROL FDA Version
Depiction in
icon bar
Mean-
ing
Depiction on
editor thread
Function and
adjustment options
?
15
STAND BY
15
Change
humid-
ity
2
h
:
30
m
40
.7
%rh
70
.6
%rh
5
.
0
%rh
5
.
0
%rh
+
-
SPWT
on off
Tolerance band
Increase
Function
Increases or decreases humidity
over a specific time to a specific
value.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration
•
Target (setpoint) humidity
•
Tolerance value above/below
•
SPWT
3
?
15
STAND BY
15
Hold
pres-
sure
+
-
Tolerance band
Alarm
Safe
on off
hold p
20
mb
10
mb
2
h
:
30
m
500
mb
Function
Maintains a specific pressure for a
specific time.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration (time or infinite ∞)
•
Pressure to be maintained
•
Tolerance value above/below
•
Alarm if limits are exceeded
•
Safe
2
?
15
STAND BY
15
Change
pres-
sure
+
-
Tolerance band
1100
mb
decrease
2
h
:
30
m
500
mb
10
mb
20
mb
SPWT
on off
Function
Increases or decreases pressure over
a specific time to a specific value.
Adjustment options
Adjustment options
•
Name of programme segment
1
•
Duration
•
Target (setpoint) pressure
•
Tolerance value above/below
•
SPWT
3
1
When run, this is displayed in the status bar of the appliance
2
When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as
specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the
actual values leaves the tolerance band, the clock timer starts again from the beginning.
3
SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint
value is reached, even if the set time has already expired. If this is "off", the programme sequence is
continued after the set time has expired, irrespective of whether the setpoint value was reached or not.