Leica Biosystems M80 Measuring User Manual
Page 10

2. Measuring pitch
5mm : 100
Stage micrometer
with 0.1mm
graduation
10
Measuring – User Manual
Calibrating
1.37mm on the stage micrometer (X)
100 intervals on the eyepiece graticule (Y)
1.37
100
mm = 0.014mm calibration value
Measuring
Intervals on the eyepiece graticule: 34
Result:
34 intervals
× 0.014mm = 0.47mm length
of measurement
distance
0
20
40
60
80
100
y
0
1
x
0
20
40
60
80
100