3.3 Setting up the instrument/Site requirements
3.2 Standard delivery
Carefully check the delivery against the packing list, delivery note and your order.
Should there be any discrepancy, please contact the Leica sales unit handling your order or
your Leica dealer.
Base unit...............................................................................................................................14 0388 35824
• Control unit with voltage selector switch, 100-120/220-240 V/50-60 Hz
• 1 screw-on tweezer holder for installation on the control unit
• 1 screw-in heatable forceps with spiral cable
• 1 set of fuses (2x T1.25 A L250 V)
• 1 set of power cables
- 1 power cable for Germany ......................................................................................14 0411 13558
- 1 power cable for UK ST/BU F-5A ............................................................................14 0411 27822
- 1 power cable for USA/Canada/Japan ...................................................................14 0411 13559
• User Manual ..................................................................................................................14 0388 83101
• Language CD ..................................................................................................................14 0388 83200
• The instrument requires an installation area of approx. 150 x 200 mm.
• The substrate must have a sufficient load capacity and rigidity with respect to the weight of
the instrument.
• Relative humidity: 20 to 80 % - non-condensing.
• Room temperature consistently between +15 °C and +40 °C.
• Elevation: up to a max. 2000 m above sea level.
• The instrument is designed for indoor use only.
• Avoid vibrations, direct sunlight and heavy variation in temperature!
• The power supply must be within the range of the length of the power cable. No extension
cable may be connected.
• The instrument MUST be connected to a grounded power socket.
Use only one of the provided power cables that is intended for the local power supply.
Do not operate the instrument in rooms with explosion hazard.
To ensure proper function of the instrument, it must be set up while maintaining a minimum
distance of 10 cm from walls and furniture.