1 product description – Labnet Z206A High Capacity, Compact Research Centrifuge User Manual
Page 8
![background image](/manuals/565098/8/background.png)
Page 8 of 24
INSTRUCTION MANUAL Z 206 A
© HERMLE Labortechnik GmbH
1 PRODUCT DESCRIPTION
1.5 Initial Setup
Before the first operation and after each rotor change, you have to select the appropriate rotor code. You
can find each rotor type in the printed order number on the rotor.
Example:
Angle
rotor
220.96
V01
≙ Rotor type “96“
Swing out rotor
220.68 V04
≙ Rotor type “68“
Switch on the unit and open the lid. Now press simultaneously the keys “lid “ (1) and “stop” (2). In the
display “rpm/rcf” then appears the selected rotor type, ex. “96”. With the adjustment knob (7) you can
select the desired rotor code. (See photo 1) To set the data in the unit please wait until the indication in
the display “rpm/rcf” disappears.
Now all important rotor parameters for the centrifuge are stored.
Photo 1
7
2
1