2 operation – Labnet Z206A High Capacity, Compact Research Centrifuge User Manual
Page 14

Page 14 of 24
INSTRUCTION MANUAL Z 206 A
© HERMLE Labortechnik GmbH
2 OPERATION
2.1.3 Overloading of rotors
The maximum load permitted for a rotor, which is determined by the manufacturer, as well as the
maximum speed allowed for this rotor (see label on rotor), must not be exceeded.
The liquids the rotors are loaded with, should have an average homogeneous density of 1,2 g/ml or less
when the rotor is running at maximum speed.
In order to spin liquids with a higher density, the speed has to be reduced according to the following
formula:
1,2
Reduced speed n
red
= x max. speed (n
max
) of the rotor
higher density
Example:
1,2
n
red
= x 4.000 = 3.360 rpm
1,7
In case of any questions, please contact the manufacturer.
2.1.4 Removing the rotor
Untighten the rotor fixing screw and lift the rotor vertically out of the centrifuge.
ATTENTION:
Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or
mechanical damage.
Do not operate with extremely corrosive substances which could damage the rotor and buckets.
2.2 Operation
2.2.1 Power switch
The power switch is located on the bottom, left side of the unit.
Photo 11
Attention:
After turning on the power switch you must open the lid of the unit first, prior starting the
centrifuge.