2 OPERATION
2.2.6 Pre-selection of brake intensity and acceleration
By pressing the key “accel/decel“ (5) this function is activated.
By pressing the key once the word “accel“ (16) flashes in the display “acc/dec“. The desired
acceleration can be pre-selected by the adjusting knob (7). The value 0 is equivalent to the lowest and
the value 9 to the highest acceleration.
By pressing the key “accel/decel“ (5) twice, in the display “acc/dec“ indicates the word “decel“ (17). Now
the desired brake intensity can be pre-selected by the adjusting knob (7). The value 9 is equivalent to
the shortest and the value 0 to longest possible brake time.
Photo 15
Acceleration and deceleration times Z 206 A(120 V / 230 V) in seconds
Acceleration
values
Deceleration
values
Rotor-Number
Level
0 Level
9 Level
0 Level
9
220.96
V01
70 40 150 35
220.97
V01
80 35 150 35
220.95
V06
60 11 30 10
220.68
V04
35 8 25 7