2 OPERATION
2.2.4 Pre-selection of speed / RCF-value
By pressing the key “rpm/rcf“ (6) the pre-selection of the speed is activated. By pressing the key once
the word “rpm“ (12) flashes.
By pressing the key once again the pre-selection of the centrifugal forces may be chosen. Flashing the
word “rcf“ (13).
You can set the desired values with the adjusting knob (7). In the display (14) the regulated value is
shown permanently, before, during and after the run.
The speed is adjustable between 200 rpm and maximum revolution of the centrifuge resp. the maximum
permissible revolution of the pre-selected rotor.
It is the same with the pre-selection of the RCF-value. The setting range is between 20 xg and the
maximum permissible centrifugal force of the rotor.
The maximum speed of the Z 206 A is 6000 rpm resp. 4180 xg.
Photo 13
Maximum speed of the rotors available for the Z 206 A
Rotor-Number Max.
Revolution
RCF
Value
220.96 V01
6000 rpm
4180 xg
220.97 V01
6000 rpm
3820 xg
220.95 V06
6000 rpm
2930 xg
220.68 V04
3500 rpm
1450 xg
Attention:
Please also check the maximum permissible speed of your test tubes.