Kds gemini 88, Specifications – KD Scientific Gemini 88 User Manual
Page 4

2
KDS Gemini 88
Specifications
Type
Microprocessor dual drive, single syringe, infuse/withdraw
Accuracy
±0.35%
Reproducibility
±0.1%
Syringe:
Type
Plastic, glass or stainless steel
Size Minimum
0.5 µl
Size Maximum
140 ml
Flow Rate:
Minimum
0.0004 µl/hr
Maximum
106.6 ml/min
Non Volatile Memory
Storage of all settings
RS-232
RJ11-4 conductor
TTL
9-pin connector
Average Linear Force
57 lbs
Drive Motor
2 motors, each 0.9° step angle motor
Motor Drive Control
Microprocessor controlled from 1/2 to 1/4
microstepping
Motor Steps per
1,600 steps at 1/2 stepping or 3,200 steps at
Rev. of Leadscrew
1/4 stepping
Step Resolution
0.33 µm/step
Step Rate:
Minimum
27.3 sec/step
Maximum
416.7 µsec/step
Pusher Travel Rate:
Minimum
0.726699 µm/min
Maximum
95.25 mm/min
Power
45 W, 1.0 A fuse
Voltage Range
95 to 130 VAC, 60 Hz; 220 to 260 VAC, 50 Hz,
selectable
Dimensions, H x W x D
15.2 x 31.1 x 28.6 cm (6 x 12.5 x 11.25 in)
Weight
6.8 kg (15 lb)
Timing Belt Drive
2:1
Lead Screw Pitch
24 threads/inch
Display
5 digits plus 14 indicator LEDS