KDS Gemini 88

External Control and Interfaces

(continued)

TTL DEVICES

The pump does not need to be configured to attach a TTL device. To attach a TTL device,

simply plug the appropriate TTL connector into the 9 pin connector on the rear of the pump.

See Appendix E for wiring specifications.

Foot Switch Input

Used to start and stop the pump. Pressing the foot switch performs the same function as

pressing the RUN/STOP key on the keypad. The foot switch connector allows remote or

automated operation of the pump.

Timer Input

Opening the timer input starts the pump. Closing the timer input stops the pump. The

timer input allows for an externally controlled pumping interval.

Pumping Direction Input

Sets the direction of pumping. Opening the directional input sets syringe 1 to infuse.

Closing the input sets syringe 1 to refill. The pumping direction input is not recognized

when data is being entered from the keypad. Syringe 2 changes direction appropriate-

ly. Also, the valve control output is set appropriately.

Directional Indicator Output

The Directional Indicator output is an indicator of the direction of pump travel. When

the output is a logical high, syringe 1 is set to infuse. A logical low indicates refill. A

valve attached to this output enables automatic selecting of infuse and refill vessels.

Running Indicator Output

Provides a signal to another device indicating whether or not the pump is running.