Kds gemini 88, Appendix c: command information – KD Scientific Gemini 88 User Manual
Page 27
25
KDS Gemini 88
Appendix C: Command Information
PUMP CHAIN COMMAND SUMMARY
All spaces in command are ignored. Maximum 5 digits per integer, or float data.
command
= >
[
adr
= >
[d]d
1 or 2 digit address
cmd
=>
RUN
Start pump
STP
Stop pump
RAT [A] [
Set/query syringe 1 rate
RAT B [
Set/query syringe 2 rate
DIA [A] [
Set/query syringe diameter 1
DIA B [
Set/query syringe diameter 2
MOD [{AUT|PRO|CON}]
Set/query pumping mode
DIR [{INF|REF|REV}]
Set/query pumping direction
PAR [{ON|OFF}]
Set/query Parallel/Reciprocal
IN
Read TTL pin d
OUT
Set TTL pin d
SAV
Save settings in NOVRAM
VER
Query software version
units
= >
{UM|UH|MM|MH}
µl/mn, µl/hr, ml/mn, ml/hr
COMMAND RESPONSE OUTPUT FORMAT:
output
= >
[
response
= >
prompt
= >
Stopped
Syringe 1 infusing
Syringe 1 withdrawing
Pump stalled
adr
= >
[d]d
1 or 2 digit pump address
message
= >
?
Unrecognized command
NA
Command not presently applicable
OOR
Entered value out of range
All spaces ignored. Maximum 5 digits per integer or floating point number.
LEGEND
float
= >
f f f f f f
f
= >
d|.
integer
= >
d d d d d
d
= >
0|1|2|3|4|5|6|7|8|9
b
= >
space
cr
= >
(ASCII 13)
lf
= >
(ASCII 10)
text
= >
(any string of ASCII characters)
< . . . >
non-terminal
upper case
terminals
[ . . . ]
optional
{ . . . }
select one
|
either-or