Kds gemini 88, Operation – KD Scientific Gemini 88 User Manual

Page 13

background image

11

KDS Gemini 88

Operation

DIAMETERS

The pump must be stopped when entering a diameter. Diameter 1 is the diameter of both

syringes, except in Proportional mode. In Proportional mode, enter the diameter of each

syringe separately; Diameter 1 for syringe 1 and Diameter 2 for syringe 2. When entering

a diameter, the 2 direction LEDs corresponding to the syringe and the “mm” LED will blink.

Enter the inside diameter of the syringe in millimeters. After a new diameter is entered the

corresponding rate is set to 0. This is done for reasons of safety. The maximum diameter is

50 mm. See Appendix A for a list of syringe diameters.

PUMPING RATES

Rate 1 is the rate of pumping of both syringes, except in Proportional mode. In Proportional

mode, enter the rate of each syringe separately, Rate 1 for syringe 1 and Rate 2 for syringe

2. When entering a rate, the directional LEDs corresponding to the syringe and the current

units LED will blink. Use the corresponding rate key (RATE 1 or RATE 2) to scroll through the

rate units: ml/min, ml/hr, µl/min and µl/hr. The minimum and maximum rates permitted

vary depending on the diameter of the syringe. If an “OOR” message, meaning “Out of

Range” is displayed when entering a rate, try using a different syringe for your application.

SELECTING THE RUN MODE

To select the run mode, press the SELECT MODE key to advance the run mode LED indicator.

Advance the run mode indicator to the desired mode, either Auto Stop, Proportional, or

Continuous mode.

Auto Stop

Both syringes pump according to DIAMETER 1 and RATE 1

until syringe 1 reaches the limit switch, pump stops.

Proportional

Each syringe pumps independently; syringe 1 according to

DIAMETER 1 and RATE 1, syringe 2 according to DIAMETER

2 and RATE 2. When syringe 1 reaches limit switch,

pump stops.

Continuous

Both syringes pump according to DIAMETER 1 and RATE 1.

When syringe 1 reaches limit switch, the direction of pumping

for both syringes is reversed and pumping continues. The

valves, if applicable, are set accordingly.

Selecting the Pumping Directions

Parallel or Reci-

Pressing the PARALLEL/ RECIPROCAL key reverses

procal Operation

pumping direction of syringe 2.

Change Direction

Pressing the CHANGE DIRECTION key reverses the pumping

directions of both syringes. When valves are used in the

system then the valve(s) setting(s) will be changed simultane-

ously with the change in direction.