KDS Gemini 88
PUMP CHAIN COMMANDS AND RESPONSES
(continued)
DIR [{INF|REF|REV}]
Request to set or query pumping direction of syringe 1.
SET:
Command
Meaning
DIR INF
Sets syringe 1 pumping direction to infusion
DIR REF
Sets syringe 1 pumping direction to refill
DIR REV
Reverses pumping direction of syringe 1
Direction of syringe 2 is set appropriately.
Query pumping direction of syringe 1:
DIR
Response will be one of:
INFUSE
REFILL
PAR [{ON|OFF}]
Request to set or query Parallel or Reciprocal setting.
SET:
Command
Meaning
PAR ON
Sets syringes to parallel pumping direction.
PAR OFF
Sets syringes to reciprocal pumping direction.
Direction of syringe 2 changes appropriately, while direction of syringe 1 does not change.
Query:
PAR
Possible Responses
Meaning
ON
Parallel
OFF
Reciprocal
IN
Request to read in the TTL logic level at the specified pin, “d”, on the external 9 pin
D-SUB connector.
Valid pin numbers for input are:
6, 7 and 8
Possible Responses
Meaning
ON
TTL level high
OFF
TTL level low
If the pin specified is invalid, a “Not Applicable” error is returned.