PUMP CHAIN COMMANDS AND RESPONSES
(continued)
DIA [{A|B}] [
Request to set or query syringe diameter settings.
Set syringe 1 diameter:
DIA diameter
DIA A diameter
Set syringe 2 diameter (only valid in proportional mode):
DIA B diameter
Corresponding syringe’s rate will be zeroed.
Diameter is of the format:
f f f f f f
Units are millimeters.
Diameter 2 is only accepted when mode is proportional. If diameter is accepted and valid,
diameter will become the new diameter. Diameter
will not
be accepted if the pump is run-
ning and a “Not Applicable” error is returned. If the diameter is invalid, an “Out Of Range”
error is returned.
Query syringe 1 diameter:
DIA
DIA A
Query syringe 2 diameter (only valid in proportional mode)
DIA B
Response is of the following format:
f f f f f f
Units are millimeters.
MOD [{AUT|PRO|CON}]
Request to set or query pumping mode
SET:
Command
Meaning
MOD AUT
Puts pump in Auto Stop mode
MOD PRO
Puts pump in Proportional mode
MOD CON
Puts pump in Continuous mode
If pump is running, command will not be accepted and a “Not Applicable” error is returned.
Query:
MOD
Possible Responses
Meaning
AUT
Auto Stop mode
PRO
Proportional mode
CON
Continuous mode