PUMP CHAIN COMMANDS AND RESPONSES

(continued)

A prompt is a string of ASCII characters sent by a pump, indicating the pump’s address and

its present state:

, 1 or 2 digit address, prompt character

Prompt Characters

Meaning

:

Pump stopped

>

Syringe 1 infusing

<

Syringe 1 refilling

*

Pump stalled

PUMP COMMANDS AND RESPONSES

RUN

Starts pumping according to the present setting of the pump. If pump already pumping, a

“Not Applicable” error is returned.

STP

Stops pump if it was running. If pump already stopped, a “Not Applicable” error is

returned.

RAT [{A|B}] [[ []]

Request to set or query infusion rate settings.

Set syringe 1 rate:

RAT rate

RAT rate units

RAT A rate

RAT A rate units

Set syringe 2 rate (only valid in proportional mode):

RAT B rate

RAT B rate units

Rate is of the format:

f f f f f f

Units are one of

Meaning

UM

µl/mn

UH

µl/hr

MM

ml/mn

MH

ml/hr

Rate 2 is only accepted when mode is Proportional. If rate is accepted and valid, rate will

become the new rate. If the rate is invalid, an “Out Of Range” error is returned.

Query syringe 1 rate:

RAT

RAT A

Query syringe 2 rate (only valid in proportional mode):

RAT B

Response is of the following format:

f f f f f f

Where units are one of the following:

ml/mn, ml/hr, µl/mn, µl/hr