Technical data
Basic type 902930/15, 902930/17 and 902930/55
Analog input
Measuring input
Pt1000 according to DIN EN 60751, in 3-wire circuit
Operating temperature range
-30 to +260 °C
Accuracy of the temperature sensing element
Class A
±(0.15 K + 0.002 × ItI) ItI = measured temperature in °C without prefix
Sensor lead resistance
≤
11 ohm per lead with 3-wire circuit
Sensor current
≤
500 µA
Measuring circuit monitoring
Detection of probe break and probe short-circuit
Output (radio transmission)
Transmitter detection (transmitter ID)
Max. 5-digit ID, factory-set, cannot be configured
Transmission interval
Adjustable via DIP switch to 5 s, 10 s, 20 s, or 45 s (factory-setting = 20 s)
Radio frequency
868.4 MHz
Transmission power
<
+10 dBm
Open air range
300 m when using the antenna wall holder for the receiver and the 3 m long antenna cable.
When installing the antenna directly on the receiver, a reduced range of approx. 40 % must be
taken into account.
Output signal
882.2 to 1977.1 ohm
־ -30 to +260 °C
Response time of the complete probe
In water 0.4 m/s / in air 3.0 m/s
Ø 4.0 mm: water t
0.5
approx. 3 s, t
0.9
approx. 7 s / air t
0.5
approx. 25 s, t
0.9
approx. 80 s
Ø 6.0 mm: water t
0.5
approx. 4 s, t
0.9
approx. 10 s / air t
0.5
approx. 32 s, t
0.9
approx. 98 s
Calibration accuracy of the electronic
components
≤
±0.05 %
1
Voltage supply
Lithium battery
Voltage: 3.6 V, rated capacity: 2.2 Ah
Operating life
Approx. 1 year with the factory-set values and at room temperature
(a fast transmission interval and a high or low ambient temperature
will reduce the operating life of the lithium battery)
Battery change
Only use the preconfigured lithium battery that is available in the accessories
(Part no. 00525539)
Environmental influences
Ambient temperature range of the case
-30 to +85 °C (case including electronic components)
If the operating temperature exceeds the max. ambient temperature, take into account the heat
input of the protection fitting into the case. For detailed information, please refer to operating
manual B902930.0.1 in the chapter "Heat input by protection tube/protection fitting“.
Storage temperature range
-40 to +85 °C (case including electronic components)
Temperature influence
≤
±0.0025 %
1
/K; per K deviation from the reference temperature 22 °C (±3 K) (case)
Resistance to climatic conditions
Rel. humidity
≤
95 % w/o condensation according to IEC 68-2-30
Vibration resistance
Max. 2 g at 10 to 2000 Hz (relating to the case including electronic components)
according to DIN IEC 60068-2-6
Admissible mechanical shock resistance
25 g/6 ms (relating to the case including electronic components)
according to DIN IEC 68-2.29 per 1000 cycles
EMC
- Interference emission
- Interference immunity
- Radio frequency spectrum
DIN EN 61326-1
Class B
Industrial requirements
ETSI EN 300 220-1 and ETSI EN 300 220-2
Case
Material
PA6 (polyamide), color: blue
Flammability class
UL 94 HB
Dimensions
Diameter approx. 32 mm, length approx. 126 mm;
For insertion length refer to order details
Protection type
IP67 according to DIN EN 60529
for basic type 902930/15 and 902930/17;
for basic type 902930/55 only with the M12 × 1 machine connector connected
Installation position
Any
Weight
Approx. 120 g (without protection fitting)
1
All accuracy specifications in % relating to the measuring range of 290 K.