Technical data – JUMO 902930 Wtrans Transmitter RTD Temperature Probe with Wireless Data Transmission Data Sheet User Manual
Page 2

JUMO GmbH & Co. KG
Delivery address: Mackenrodtstraße 14
36039 Fulda, Germany
Postal address:
36035 Fulda, Germany
Phone:
+49 661 6003-0
Fax:
+49 661 6003-607
E-mail:
Internet:
www.jumo.net
JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:
+44 1279 635262
E-mail:
Internet: www.jumo.co.uk
JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone: 315-697-5866
1-800-554-JUMO
Fax:
315-697-5867
E-mail:
Internet: www.jumo.us
2012-03-02/00473251
Data sheet 902930
Page 2/25
Technical data
Basic type 902930/10, 902930/12, 902930/20, 902930/22, 902930/50, and 902930/60
Analog input
Measuring input
Pt1000 according to DIN EN 60751, in 3-wire circuit
Operating temperature range
-30 to +260 °C for basic type 902930/10 and 902930/20;
-50 to +600 °C for basic type 902930/12 and 902930/22;
-200 to +600 °C for basic type 902930/50 and 902930/60;
Observe the admissible ambient temperature range for the case.
Accuracy of the temperature sensing element
Class A
±(0.15 K + 0.002 × ItI) ItI = measured temperature in °C without prefix
Sensor lead resistance
≤
11 ohm per lead with 3-wire circuit
Probe current
≤
500 µA
Measuring circuit monitoring
Detection of probe break and probe short circuit.
Output (radio transmission)
Transmitter detection (transmitter ID)
Max. 5-digit ID, factory-set, can be configured customer-specific
Transmission interval
Adjustable from 1 to 3600 s:
for basic type 902930/10, 902930/12, and 902930/50 (factory setting = 10 s);
Adjustable from 5 to 3600 s:
for basic type 902930/20, 902930/22, and 902930/60 (factory setting = 15 s)
Radio frequency
868.4 MHz (Europe);
915 MHz (America, Australia, Canada, and New Zealand);
Ten frequencies can be set in the 915 MHz frequency band.
Transmission power
<
+10 dBm
Open air range
300 m for 868.4 MHz, 100 m for 915 MHz
and when using the antenna wall holder for the receiver and the 3 m long antenna cable. When
installing the antenna directly on the receiver, a reduced range of approx. 40 % must be taken
into account.
Output signal
882.2 to 1977.1 ohm
־ -30 to +260 °C; 185.2 to 3137.1 ohm ־ -200 to +600 °C;
Response time of the complete probe
In water 0.4 m/s / in air 3.0 m/s
Ø 4.0 mm: water t
0.5
approx. 3 s, t
0.9
approx. 7 s / air t
0.5
approx. 25 s, t
0.9
approx. 80 s
Ø 6.0 mm: water t
0.5
approx. 4 s, t
0.9
approx. 10 s / air t
0.5
approx. 32 s, t
0.9
approx. 98 s
Calibration accuracy of the electronic
components
≤
±0.05 %
1
Configuration
With setup program
Configurable parameters
Transmitter detection (max. 5-digit ID), transmission interval, radio frequency (only at 915 MHz)
Voltage supply
Lithium battery
Voltage: 3.6 V, rated capacity: 2.2 Ah
for basic type 902930/10, 902930/12, and 902930/50;
Voltage: 3.6 V, rated capacity: 1.7 Ah
for basic type 902930/20, 902930/22, and 902930/60
Operating life
Approx. 1 year with the factory-set values and at room temperature
(a fast transmission interval and a high or low ambient temperature
will reduce the operating life of the lithium battery)
Battery change
Only use the preconfigured lithium battery that is available in the accessories
Environmental influences
Ambient temperature range of the case
-30 to +85 °C (case including electronic components)
for basic type 902930/10, 902930/12, and 902930/50;
-25 to +125 °C (case including electronic components)
for basic type 902930/20, 902930/22, and 902930/60
Storage temperature range
-40 to +85 °C (case including electronic components)
for basic type 902930/10, 902930/12, and 902930/50;
-40 to +125 °C (case including electronic components)
for basic type 902930/20, 902930/22, and 902930/60
Temperature influence
≤
±0.0025 %
1
/K; per K deviation from the reference temperature 22 °C (±3 K) (case)
Resistance to climatic conditions
Rel. humidity
≤
95 % w/o condensation according to IEC 68-2-30
Vibration resistance
Max. 2 g at 10 to 2000 Hz (relating to the case including electronic components)
according to DIN IEC 60068-2-6
Admissible mechanical shock resistance
25 g/6 ms (relating to the case including electronic components)
according to DIN IEC 68-2.29 per 1000 cycles
1