beautypg.com

JUMO 707060 Wtrans B - Programmable Head Transmitter with Radio Transmission Data Sheet User Manual

Page 5

background image

2011-09-30/00532898

Data Sheet 707060

Seite 5/10

JUMO GmbH & Co. KG
Delivery address:Mackenrodtstraße 14,

36039 Fulda, Germany

Postal address: 36035 Fulda, Germany
Phone:

+49 661 6003-0

Fax:

+49 661 6003-607

E-mail:

[email protected]

Internet:

www.jumo.net

JUMO Instrument Co. Ltd.
JUMO House
Temple Bank, Riverway
Harlow, Essex CM20 2DY, UK
Phone: +44 1279 635533
Fax:

+44 1279 635262

E-mail:

[email protected]

Internet: www.jumo.co.uk

JUMO Process Control, Inc.
8 Technology Boulevard
Canastota, NY 13032, USA
Phone:

315-697-5866
1-800-554-JUMO

Fax:

315-697-5867

E-mail:

[email protected]

Internet: www.jumo.us

Output (radio transmission)

Electrical data

Environmental influences

Transmitter in the B head with antenna-battery case

Probe ID

Max. 5-digit ID, factory-set, can be configured to customer-request

Transmission interval

Adjustable from 1 to 3600 s (factory setting: 15 s)

Transmission frequency

868.4 MHz (Europe)

Transmission power

+10 dBm

Open air range

Max. 300 m when using the antenna wall-holder of the receiver and the 3 m antenna
line. When installing the antenna directly on the receiver, a reduced range of approx.
40 % must be taken into account.

Output signal

Thermocouple

Voltage (mV)

RTD temperature probe

Resistance (

Ω

)

Resistance / potentiometer

Percent (%) or resistance (

Ω

)

Resistance transmitter

Percent (%)

Voltage

Voltage (mV)

Configuration

With setup program

Configurable parameters

Probe ID (max. 5-digit ID) and transmission interval

Voltage supply

Lithium battery

Rated voltage: 3.6 V, rated capacity: 2.2 Ah; size AA

Operating life

Approx. 1 year with the factory-set values (transmission interval = 15 s) and at room
temperature (brief transmission interval and high or low ambient temperature reduce
the battery operating life)

Battery change

Only use the lithium battery available as an accessory

Ambient temperature range

-30 to +85 °C

Storage temperature range / storage humidity

-40 to +85 °C / relative humidity

95 %

Temperature influence

a

a

All accuracy values in % refer to the maximum measuring range.

Thermocouple

± 0.005 % / K deviation from 22 °C plus accuracy of the cold junction

RTD temperature probe

± 0.005 % / K deviation from 22 °C

Resistance / potentiometer

± 0.01 % / K deviation from 22 °C

Resistance transmitter

± 0.01 % / K deviation from 22 °C

Voltage

± 0.005 % / K deviation from 22 °C

Climate class

10 cycles at 10 °C / 80 °C, according to IEC 68-2-30,
relative humidity 95 %, during operation

Vibration resistance

According to GL characteristic line 2

Admissible mechanical shock resistance

10 g / 6 ms, according to DIN IEC 68-2.29

Electromagnetic compatibility (EMC)

According to DIN EN 61326-1

Interference emission

Class B

Interference immunity

Industrial requirements

Radio frequency spectrum

ETSI EN 300 220-1 (V 2.3.1) and ETSI EN 300 220-2 (V 2.3.1)