Technical data
Analog input
RTD temperature probe
Thermocouples
Designation
Standard
Measuring range
Measuring accuracy
Pt100
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt500
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt1000
DIN EN 60751
-100 to +200 °C
± 0.1 K
(TK value = 3.85 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Ni 100
DIN 43760
-60 to +250 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Ni 500
DIN 43760
-60 to +150 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Ni 1000
DIN 43760
-60 to +150 °C
±
0.2 K
(TK value = 6.18 × 10
-3
1/K)
Pt100
JIS 1604
-100 to +200 °C
± 0.1 K
(TK value = 3.917 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Pt50
ST RGW 1057 1985
-200 to +600 °C
±
0.2 K
(TK value = 3.91 × 10
-3
1/K)
Pt100
GOST 6651-94 A.1
-100 to +200 °C
± 0.1 K
(TK value = 3.91 × 10
-3
1/K)
-200 to +600 °C
±
0.2 K
Cu50
GOST 6651-94 A.4
-50 to +200 °C
±
0.2 K
(TK value = 4.26 × 10
-3
1/K)
Cu100
GOST 6651-94 A.4
-50 to +200 °C
±
0.2 K
(TK value = 4.26 × 10
-3
1/K)
Connection type
2-wire or 3-wire circuit
Sensor current
< 0.5 mA
Lead compensation
Not required for 3-wire circuit (max. admissible 11
Ω
per line).
Max. configurable lead wire resistance for 2-wire circuit: 22
Ω
.
Designation
Standard
Measuring range
Measuring accuracy
a
Fe-CuNi „L“
DIN 43710
-200 to +900 °C
± 0.1 %
Fe-CuNi „J“
DIN EN 60584
-210 to +1200 °C
± 0.1 % as of -100 °C
Cu-CuNi „U“
DIN 43710
-200 to +600 °C
± 0.1 % as of -100 °C
Cu-CuNi T
DIN EN 60584
-270 to +400 °C
± 0.1 % as of -150 °C
NiCr-Ni K
DIN EN 60584
-270 to +1372 °C
± 0.1 % as of -80 °C
NiCr-CuNi „E“
DIN EN 60584
-270 to +1000 °C
± 0.1 % as of -80 °C
NiCrSi-NiSi „N“
DIN EN 60584
-270 to +1300 °C
± 0.1 % as of -80 °C
Pt10Rh-Pt „S“
DIN EN 60584
-50 to +1768 °C
± 0.15 % as of 20 °C
Pt13Rh-Pt „R“
DIN EN 60584
-50 to +1768 °C
± 0.15 % as of 50 °C
Pt30Rh-Pt6Rh „B“
DIN EN 60584
0 to 1820 °C
± 0.15 % as of 400 °C
W5Re-W26Re „C“
0 to 2320 °C
± 0.15 %
W3Re-W25Re „D“
0 to 2495 °C
± 0.25 %
W3Re-W26Re
0 to 2400 °C
± 0.15 %
Chromel-Copel
-200 to +800 °C
± 0.1 % as of -80 °C
Chromel-Alumel
-200 to +1372 °C
± 0.1 % as of -80 °C
PLII (Platinel II)
0 to 1395 °C
± 0.15 %