12.4. Status messages

Status messages

Description

00 MANUAL STOP

Circulator in „OFF“ state.

01 MANUAL START

Circulator in keypad control mode.

02 REMOTE STOP

Circulator in „r OFF“ state.

03 REMOTE START

Circulator in remote control mode.

12.5. Error messages

Error messages

Description

-01 LOW LEVEL ALARM

Low liquid level alarm.

-02 REFRIGERATOR ALARM

Control cable of the refrigerated circulator or MVS
solenoid valve controller short-circuited or interrupted.

-03 EXCESS TEMPERATURE

WARNING

High temperature warning.

-04 LOW TEMPERATURE WARNING

Low temperature warning.

-05 WORKING SENSOR ALARM

Working temperature sensor short-circuited or interrupted.

-06 SENSOR DIFFERENCE ALARM

Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 35 K

-07 I

2

C-BUS ERROR

Internal error when reading or writing the I

2

C bus.

-08 INVALID COMMAND

Invalid command.

-09 COMMAND NOT ALLOWED IN

CURRENT OPERATING MODE

Invalid command in current operating mode.

-10 VALUE TOO SMALL

Entered value too small.

-11 VALUE TOO LARGE

Entered value too large.

-12 TEMPERATURE MEASUREMENT

ALARM

Error in A/D converter.

-13 WARNING : VALUE EXCEEDS

TEMPERATURE LIMITS

Value lies outside the adjusted range for the high and low
temperature warning limits. But value is stored.

-14 EXCESS TEMPERATURE

PROTECTOR ALARM

Excess temperature protector alarm

-15 EXTERNAL SENSOR ALARM

External control selected, but external Pt100 sensor not
connected.

-20 WARNING: CLEAN CONDENSOR

OR CHECK COOLING WATER
CIRCUIT OF REFRIGERATOR

Cooling of the condenser is affected.
Clean air-cooled condenser.
Check the flow rate and cooling water temperature on
water-cooled condenser.

-21 WARNING: COMPRESSOR STAGE

1 DOES NOT WORK

Compressor stage 1 does not work.

-22 WARNING: COMPRESSOR STAGE

2 DOES NOT WORK

Compressor stage 2 does not work.