Technical specifications – JULABO SE-Z Heating Circulators User Manual
Page 10

Operator responsibility – Safety recommendations
10
2.4.
Technical specifications
Heating circulators
HE-4
SE-6
SE-12
SE-26
Working temperature range
°C
20 … 250
20 ... 300
20 ... 300
20 ... 300
Temperature stability
°C
±0.01
±0.01
±0.01
±0.01
Bath opening (WxL)
cm
13x15
13x15
22x15
22x30
Bath depth
cm
15
20
20
20
Filling volume
liters
3 ... 4,5
4,5 ... 6
7,5 ... 12
18 ... 26
Overall dimensions (WxDxH)
cm
21x42x40
21x43x44
30x43x47
36x61x47
Weight
kg
11
13,5
14,0
27,0
Bridge mounted circulator
SE-Z
Working temperature range
°C
20 ... 300
Temperature stability
°C
±0.01
Immersion depth max./min.
cm
18/9
Overall dimensions (WxDxH)
cm
32x17x33
Weight
kg
7
HE
SE
Mains power connection
230 V/50-60 Hz
V/Hz
190-253 / 50-60
190-253 / 50-60
Current draw (at 208 V / 230 V)
A
8 / 9
12 / 13
Mains power connection
115 V/50-60 Hz
V/ Hz
90-127 / 50-60
Current draw (at 100 V / 115 V)
A
9 / 10
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
bath fluid: water
operating temperature: 70 °C
Technical changes without prior notification reserved.
HE
SE
Temperature selection
digital
via keypad
indication on VFD COMFORT-DISPLAY
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY
Resolution
°C
0.01
0.01
ATC3
INT/EXT
°C
±3 / ±9
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2,0
3,0
Heater wattage (at 115 V)
kW
1,0
---
Electronically adj. pump capacity
stages
1 ... 4
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
22 ... 26
Max. pressure at 0 liters
bar
0.7
0.7
Max. suction at 0 liters
bar
0.4
0.4
Electrical connections:
See page 11
Ambient temperature
°C
5 … 40
5 ... 40